AC/DC Hystograph HG 200
AC |
|
AC / DC Hystograph |
|
|
|
Measurement of all magnetic materials
(hard magnetic AlNiCo, ferrite, SmCo, NdFeB, plastic composite; soft magnetic:
electrical steel, ferrite …etc.) |
|
Measurement with constant flux
alteration d/dt or
sinusoidal polarization |
|
Measurement with J-compensated
surrounding coils, pole or field coils or strip measuring coil, ring core
sensor, punched part sensor, Epstein frame, sheet measuring coil |
|
Measurement at higher temperatures up to
200°C |
|
Fully computer-controlled measuring
system |
|
Real time display of hysteresis during
the measuring process |
|
MAG Expert software for measurement,
presentation and integration into QM systems |
|
Windows
desktop |
|
Determination of the characteristic properties of all magnetic materials.
Measuring process according to IEC 60404 ff. or to customer
specifications.
Highly accurate measurements with very good reproducibility.
Process-controlled monitoring and regulation of the increase in current.
When measuring hard magnetic materials the process runs with constant flux
alteration d/dt and
avoids interference caused by eddy currents and phase displacement between the
field strength and polarization measurement.
With soft magnetic materials the measurement takes place with sinusoidal
induction. Processing of the measured values with MAG/MPG Expert software for
measurement, display and integration into QM systems.
Consists of measuring table with power supply unit and measuring electronics,
electromagnet, Epstein frame, measuring system and computer hardware with
MAG/MPG Expert evaluation software.
For measuring permanent magnets: J-compensated coils with a thickness of 1mm
and diameters of 10 – 40 mm, pole measuring coils with diameters of 3 – 9 mm, as
well as special sizes and shapes.
For measuring soft magnetic materials: SST measuring sensors for the quality
control of strips, sheets, slit strips in all manufacturing widths; no complex
sample preparation required.
Punched part sensor for nondestructive testing of punched parts in all
geometries, e.g. when testing punched parts for motors. Ring core testing module
for testing the magnetic properties of ring cores, stators or transformers
directly on the component under manufacturing conditions.
Technical Data : AC/DC Hystograph HG 200
AC
|
Measuring fixture: |
2 free of drift fluxmeters F
01 |
Measuring coils: |
surrounding coils, pole coils, field
coils |
Power supply: |
0 to ± 125 V, 0 to ± 25 A (50 A
optional) |
Electromagnet: |
interchangeable with a maximum diameter
of 92 mm |
Measuring poles: |
< ± 1 µVs/minute |
Computer: |
PC, monitor, printer |
Software: |
MAG/MPG Expert under Windows |
Cabinet: |
measuring table with container |
Dimensions: |
760 x 1760 x 800 mm (height X width X
depth) |
Mains supply: |
3 x 200 – 3 x 400 V AC, 16 A 50 / 60 Hz | | |
|
|
|
|
|
|